Uni-Logo
Sie sind hier: Startseite Laboratories Woias, Peter Projects

Projects

Aktive Microports

Projektbild

Project description

A miniaturized autonomous drug delivery system to carry chemotherapeutic studies in rats has been developed and tested (Fig. 1). The heart of the dosing system is a backpressure independent piezoelectric micropump. A back pressure independent operation up to 30 kPa was demonstrated with flow rates below 1.5 µl / min. The overall system has dimensions of 49 x 36 x 26 mm³ and a curb weight of 36g with a reservoir size of 10 ml and can be operated via an infrared interface. The system has been successfully used in a study of metronomic therapy with doxorubicin prostate carcinomas .

Start/End of project

01.01.2005 until 31.12.2007

Project manager

Frank Goldschmidtboeing (Prof. Dr. Peter Woias)

Contact person

Frank Goldschmidtboeing
Phone:0761 / 203-7496

Partners

Klinik für Tumorbiologie, Prof. Dr. rer. nat Ulrich Massing